D-Link DSA-3100P - B/W Thermal Line Printer Uživatelský manuál Strana 29

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 32
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 28
29
Artis zee
Floor-mounted system for surgical angiography
Ambient conditions (operation)
Examination and control room Temperature range: + 15°C to + 30°C (recommended temp. 22°C [72°F])
Relative humidity: 20 – 75% below dew point
Imaging system Temperature range: + 10°C to + 35°C
Relative humidity: 20 – 75% below dew point
Air flow: 630 m
3
/h
Noise level: < 53 dB (A)
Generator Temperature range: + 10°C to + 3C
Relative humidity: 20 – 75% below dew point
Temperature gradient: max. 5°C/h
Air flow: 160 m
3
/h
Noise level: < 55 dB (A)
Cooling system Cooling air: + 5°C to + 3C
(for MEGALIX tube) Air flow: 1100 m
3
/h
Noise level: 55 dB (A) at 50 Hz;
59 dB (A) at 60 Hz
System control cabinet 1 Temperature range: + 15°C to + 3C
Relative humidity: 20 – 75% below dew point
Temperature gradient: max. 5°C/h
Air flow: 500 m
3
/h
Noise level: 48 dB (A)
System control cabinet 2 Temperature range: + 10°C to + 35°C
(only for OR table) Relative humidity: 20 – 75% below dew point
Temperature gradient: max. 5°C/h
Air flow: n/a
Noise level: n/a
Zobrazit stránku 28
1 2 ... 24 25 26 27 28 29 30 31 32

Komentáře k této Příručce

Žádné komentáře