9
COMMAND LIST DELTA
COMMAND SYNTAX LIST D
ARGUMENTS None
DESCRIPTION Lists the 16 delta zero correction values. These are used in the conversion
of raw counts to Engineering Units(EU).
EXAMPLE To view the current delta zero correction values:
Type: LIST D<CR>
The DSA will return the delta zero correction values. They could appear as
follows.
SET DELTA0 0
SET DELTA1 1
SET DELTA2 0
SET DELTA3 1
SET DELTA4 0
SET DELTA5 1
SET DELTA6 0
SET DELTA7 0
SET DELTA8 1
SET DELTA9 0
SET DELTA10 1
SET DELTA11 0
SET DELTA12 1
SET DELTA13 0
SET DELTA14 1
SET DELTA15 0
NOTE The delta values will vary as the module ages. These values are the
difference between the current zero offset value obtained in the most recent
CALZ and the zero value in the coefficient table.
COMMAND DIGITAL OUTPUT
COMMAND SYNTAX DOUT <Number><Value>
ARGUMENTS Number The digital output channel 0 through 7
Value 1 = On
0 = Off
DESCRIPTION Commands the DSA to energize or de-energize a Digital Output Channel.
RETURNS <nl>
<nl> - end of line.
EXAMPLE 1 To energize Digital Output 1:
Type: DOUT 1 1<CR>
EXAMPLE 2 To de-energize Digital Output 6:
Type: DOUT 6 0<CR>